Connection
First, connect the cable to the pH sensor and then connect the other end of the cable to the first slot on the universal fixture. Make sure that no liquid splashes on the electronics during assembly or during measurement - the measuring probe may come into contact with the liquid, but the electronic part of the module must remain dry.
When connecting, the universal tool must be turned off and the micro:bit must be disconnected from the power supply. Once the sensor is properly connected to the universal tool, you can turn on the device using the power switch and start the pH measurement.
Program
Program for probe V1
For the sensors to function properly, calibration is necessary. The procedure for performing the calibration can be found at manufacturer's website or in the enclosed instructions.
Program for probe V2
For the sensors to function properly, calibration is necessary. The procedure for performing the calibration can be found at manufacturer's website or in the enclosed instructions.
In conjunction with the microbit, the V1 probe can only be used for measurements up to pH < 12 (12 is the maximum value). Exceeding this limit may damage the microbit.



